===== ValiFrame USB4 Module ===== ======================================================================================================================= 1.2 [released 2020-11-03] ======================================================================================================================= 1.2.1 [available 2021-01-13] supported M8041A/M8062A, M8070B FW rev. 7.2.40.2 supported M8045A/M8054A, M8070B FW rev. 7.2.40.2 supported DSOX, FW rev. 6.55.00901 supported UXR1102A, FW rev. 10.25.00702 bugfix Compliance test with USB4 ETT stopped with an critical error although the DUT's result was ok known issue USB4 ETT implementation is in beta state and may not work as expected known issue Testing TBT automated scripts: TenLira 4.10 doesn't support the 'uc_Reset.tbc' script! 1.2.0 [available 2020-11-03] tested with M8041A/M8062A, M8070B FW rev. 7.2.40.2 tested with M8045A/M8054A, M8070B FW rev. 7.2.40.2 tested with DSOX, FW rev. 6.55.00901 tested with UXR1102A, FW rev. 10.25.00702 new Support of USB4 CTS 1.0 new Testing of up to 10 ports of a DUT new Additional test procedures to check the USB4 ETT installation and environment new First implementation of USB4 ETT support (beta!) new TestToolServer to support the USB4 ETT on a remote PC is included in the Valiframe installer modification Always use random jitter bandwidth of 500MHz (according to CTS 1.0) modification Multi error-burst test now runs on M8040A for all kind of DUTs modification Configuration of the ISI source (physical trace or internal M8020A-emulation) is now done in the Configure-DUT dialog bugfix Swapped fixtures in the connection diagram for insertion loss calibration bugfix Closing the Valiframe main window was handled in a confusing and unexpected way bugfix Jitter tolerance characterization did not iterate correctly in some rare corner cases known issue USB4 ETT implementation is in beta state and may not work as expected known issue Testing TBT automated scripts: TenLira 4.10 doesn't support the 'uc_Reset.tbc' script! ======================================================================================================================= 1.1 [released 2020-08-24] ======================================================================================================================= 1.1.2 [available 2020-10-12] modification Allow to select different schemes of DUT-Reset (to support of different TenLira versions) bugfix The initial Tx preset was not always set correct in TP3' tests tested with TenLira 3.16 (recommended) tested with TenLira 4.10 known issue Testing TBT automated scripts: TenLira 4.10 doesn't support the 'uc_Reset.tbc' script! 1.1.1 [available 2020-09-29] Information Updated CodeMeter Runtime to version 7.10a to fix vulnerabilities CVE-2020-14513, CVE-2020-14519, CVE-2020-14509, CVE-2020-14517, CVE-2020-16233, and CVE-2020-14515 bugfix Swapped fixtures in the connection diagram for insertion loss calibration 1.1.0 [available 2020-08-24] tested with M8041A/M8062A, M8070B FW rev. 7.2.40.2 tested with M8045A/M8054A, M8070B FW rev. 7.2.40.2 tested with M8040A, M8070B FW rev. 7.2.40.2 tested with DSOX, FW rev. 6.55.00702 tested with DSA-Z, FW rev. 6.55.00702 tested with UXR1102A, FW rev. 10.25.00702 bugfix The initial PRBS amplitude of M8040 sometimes was too low in setups with PickOffTees and splitter bugfix Assigned Tx preset at the TP3-EyeCalibration was not always correct bugfix Assigned Jitter was not always correct bugfix Separate CTLE calibration needed for each jitter frequency bugfix Due to rounding, DFE sometimes was configured to values slightly above 50 mV bugfix Use the DFE settings found during CTLE calibration for TP3 eye calibration instead of reconfigure DFE at each measurement bugfix CTLE configuration on the scope must be "Equalize in place", not "Display as function" bugfix Check for monotony at EyeHeight calibration at TP3 was not working correct bugfix Don't show ISI properties in TP3' calibrations bugfix Don't query the chip type in manual USB4 tests (there are no predefined chip types for USB4) bugfix Jitter tolerance characterization sometimes exceeded the allowed PJ limits of the BERT bugfix Confusing description in the connection diagram of TP3 test procedures bugfix The query for max PJ in the MultiErrorBurstTest sometimes returned wrong and too low value bugfix If running tests via TestScriptServer, sometimes the tempfile could not be accessed by the TclScript new Added step to evaluate the intrinsic RJ of the scope new Added step to calibrate the channel insertion loss at TP3 (required ISI trace or internal ISI) new Using Infiniium eye height measurement method instead of eye mask measurements new Added summary tables to each calibration test point new Handling of outliers measurement results in the input eye calibration at TP3 modification Make external ISI as default settings when using M8020A system modification Use pure instrument set values in cal tables and graphs modification Improved the ordering and categorizing of parameters in the reports modification Increased the step size in AC CommonMode calibration to have better monotonic behavior modification Smaller step size for the final Input Eye Calibration at TP3 to have better result distribution modification Better columns / x-axis name for signal amplitudes modification Update instrument images modification Repeat the Eye width calibration, before it would fail because of EyeHeight out of spec limits modification Don't report dropped eye measurements in the the final Input Eye Width calibration modification Adding extra wait statement after the data acquisition step. This gives more stable results and significant bigger eyes modification Repeat the DFE AutoDelay several times to avoid uncompleted DFE configuration modification Use the new defined Tx preset 14 modification Timing improvements in the eye height measurement at TP3' modification Don't complain about USB connections to the scope, it's as fast as hislip modification Don't require adjustable ISI license for the transition time adaption known issue Error Burst Test may not work for every DUT on M8040A known issue FrequencyVariation test not supported by M8040A system (not enough PJ capability of M8045A module) known issue Captive devices not included in v1.1 because of outdated CTS known issue PDO testing is not supported due to missing functionality in Wilder TR controller API ======================================================================================================================= 1.0 [released 2020-04-17] ======================================================================================================================= 1.0.1 [available 2020-05-26] modification More prominent indication of the used fixture type in the connection diagrams modification Explicit call for tclsh85 instead of calling tclsh to avoid the usage of wrong tcl versions bugfix iTBT DUTs were not recognized correctly, when testing with TBT proprietary scripts 1.0.0 [available since 2020-04-17] tested with J-BERT M8045A+M8196A and firmware M8070B FW rev. 6.7.500.6 tested with J-BERT M8041A+M8062A and firmware M8070B FW rev. 6.7.500.6 tested with UXR0134A FW rev. 10.12.05302 tested with DSA-Z FW rev. 6.50.01004 tested with DSA-V FW rev. 6.50.01004 tested with 81160A, FW 2.0.0.0-2.6 tested with N5181A MXG, FW A.01.50 tested with Legacy controller and TBT electrical scripts 0.8.9.1 new Adapted to USB4 CTS 0.96 (Jan 30, 2020) new Changed to BLM (BitifEye License Manager) licensing known issue RJ LowPass Bandwidth of J-BERT M8041A+M8062A at 10.0G and 10.3125G is limited to 500MHz known issue FrequencyVariation test not supported by M8040A system (not enough PJ capability of M8045A module) known issue Captive devices not included in v1.0 because of outdated CTS known issue PDO testing is not supported due to missing functionality in Wilder TR controller API known issue Correlation of UXR Sr with DSO-V/Z is not approved